Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 21: (1 year) CAL VER /5520 INSTRUMENT: Fluke 23: (1 year) CAL VER /5520 INSTRUMENT: Fluke 75: (1 year) CAL VER /5520 INSTRUMENT: Fluke 77: (1 year) CAL VER /5520 DATE: 26-May-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 33 NUMBER OF LINES: 172 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034 Jan. 1984). # # Compatibility: # 5500/CAL 5.1 or MET/CAL 5.1 or later # # Subprocedures: # None # # Required Files: # 7x_dsp.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Allow the UUT to stablize to room temperature 1.005 DISP 23degC +/-5degC (73degF +/-9degF). 1.005 DISP 1.005 DISP Check the fuses and battery, and replace them if 1.005 DISP necessary. 1.006 HEAD {DISPLAY TEST} 1.007 PICE 7x_DSP 2.001 HEAD {RESISTANCE TESTS} 2.002 DISP Select the Ohms Function. 2.003 DISP Connect the 5520 to the UUT as follows: 2.003 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 2.003 DISP [32] 5520A NORMAL LO to UUT COM 2.003 DISP [32] 5520A AUX HI to UUT V/Ohm/Diode 2.003 DISP [32] 5520A AUX LO to UUT COM 2.004 JMPT 8.003 Fluke 23: (1 year) CAL VER /5520 2.005 JMPT 8.003 Fluke 77: (1 year) CAL VER /5520 # # 75/21 2.006 TARGET 2.007 HEAD RESISTANCE TESTS: {320 Ohm Range} 2.008 5520 320 100.0Z 0.9U CW 3.001 HEAD RESISTANCE TESTS: {3200 Ohm Range} 3.002 5520 3200 1000Z 8U CW 4.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 4.002 5520 32 10.00kZ 0.08U CW 5.001 HEAD RESISTANCE TESTS: {320 kOhmRange} 5.002 DISP Connect the 5520 to the UUT as follows: 5.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 5.002 DISP [32] 5520A NORMAL LO to UUT COM 5.003 5520 320 100.0kZ 0.8U 2W 6.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 6.002 5520 3.2 1.000MZ 0.008U 2W 7.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 7.002 5520 32 10.00MZ 0.26U 2W 8.001 JMP 14.001 # # 77/23 8.002 TARGET 8.003 HEAD RESISTANCE TESTS: {320 Ohm Range} 8.004 5520 320 100.0Z 0.7U CW 9.001 HEAD RESISTANCE TESTS: {3200 Ohm Range} 9.002 5520 3200 1000Z 6U CW 10.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 10.002 5520 32 10.00kZ 0.06U CW 11.001 HEAD RESISTANCE TESTS: {320 kOhmRange} 11.002 DISP Connect the 5520 to the UUT as follows: 11.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 11.002 DISP [32] 5520A NORMAL LO to UUT COM 11.003 5520 320 100.0kZ 0.6U 2W 12.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 12.002 5520 3.2 1.000MZ 0.006U 2W 13.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 13.002 5520 32 10.00MZ 0.21U 2W 14.001 HEAD {DIODE TEST} 14.002 M5520 RNGLK 3.3V 14.003 DISP Select the Diode/Audible-Tone Function 14.004 5520 0.090V S 2W 14.005 EVAL -s "Beeper On" : Is the beeper on? 15.001 5520 0.110V S 2W 15.002 EVAL -s "Beeper Off" : Is the beeper off? 16.001 M5520 * 16.002 HEAD {DC MILLIVOLT TEST} 16.003 DISP Select the 300mV DC function. 16.004 JMPT 17.003 Fluke 23: (1 year) CAL VER /5520 16.005 JMPT 17.003 Fluke 77: (1 year) CAL VER /5520 # # 75/21 16.006 TARGET 16.007 HEAD DC MILLIVOLT TEST: {320mV Range} 16.008 5520 320 300.0mV 1.6U 2W 17.001 JMP 18.001 # # 77/23 17.002 TARGET 17.003 HEAD DC MILLIVOLT TEST: {320mV Range} 17.004 5520 320 300.0mV 1.0U 2W 18.001 HEAD {DC VOLTAGE TESTS} 18.002 DISP Select the V DC function. 18.003 JMPT 22.003 Fluke 23: (1 year) CAL VER /5520 18.004 JMPT 22.003 Fluke 77: (1 year) CAL VER /5520 # # 75/21 18.005 TARGET 18.006 HEAD DC VOLTAGE TESTS: {3.2V Range} 18.007 5520 3.2 2.700V 0.015U 2W 19.001 HEAD DC VOLTAGE TESTS: {32V Range} 19.002 5520 32 27.00V 0.15U 2W 20.001 HEAD DC VOLTAGE TESTS: {320V Range} 20.002 5520 320 270.0V 1.5U 2W 21.001 HEAD DC VOLTAGE TESTS: {1000V Range} 21.002 5520 1000 1000V 7U 2W 22.001 JMP 26.001 # # 77/23 22.002 TARGET 22.003 HEAD DC VOLTAGE TESTS: {3.2V Range} 22.004 5520 3.2 2.700V 0.009U 2W 23.001 HEAD DC VOLTAGE TESTS: {32V Range} 23.002 5520 32 27.00V 0.09U 2W 24.001 HEAD DC VOLTAGE TESTS: {320V Range} 24.002 5520 320 270.0V 0.9U 2W 25.001 HEAD DC VOLTAGE TESTS: {1000V Range} 25.002 5520 1000 1000V 5U 2W 26.001 HEAD {AC VOLTAGE TESTS} 26.002 DISP Select the V AC function. 26.003 HEAD AC VOLTAGE TESTS: {3.2V Range} 26.004 5520 3.2 2.700V 0.056U 100H SI 2W 27.001 5520 3.2 2.700V 0.056U 500H SI 2W 28.001 HEAD AC VOLTAGE TESTS: {750V Range} 28.002 5520 750 750V 17U 100H SI 2W 29.001 5520 750 750V 17U 1000H SI 2W 30.001 HEAD {LOW CURRENT TESTS} 30.002 DISP Select the A DC function. 30.003 DISP Connect the 5520 to the UUT as follows: 30.003 DISP [32] 5520A AUX HI to UUT 300mA 30.003 DISP [32] 5520A AUX LO to UUT COM 30.004 HEAD LOW CURRENT TESTS: {32mA Range} 30.005 5520 32 27.00mA 0.43U 2W 31.001 HEAD LOW CURRENT TESTS: {320mA Range} 31.002 5520 320 200.0mA 4.2U 2W 32.001 JMPT 33.001 Fluke 21: (1 year) CAL VER /5520 32.002 HEAD {HIGH CURRENT TEST} 32.003 DISP Connect the 5520 to the UUT as follows: 32.003 DISP [32] 5520A 20A to UUT 10A 32.003 DISP [32] 5520A AUX LO to UUT COM 32.004 HEAD HIGH CURRENT TEST: {10A Range} 32.005 5520 10 2.00A 0.06U 2W 33.001 END